Pacemaker Analyzers
External Pacemaker Analyzer - EXPMT 2000
Product Code : 01
EXPMT 2000 External Pacemaker Analyzer is FDA approved and has evolved from the best selling EXPMT 100 Pacer Analyzer. It will test and verify all the functions of any External Pacemaker including transvenous, transcutanious (transthoracic), and Atrial and Ventricular sequential pacemakers.
Operation of the analyzer is menu driven via eight tactile keys and a large four line LCD display. Test measurements are simultaneously presented on the display. Five ECG snaps are standard along with an RS-232 serial port that allows for printing both a heading and test measurements on a separate printer. Information may also be sent to an external device. The external test load has twelve user-selectable loads from 100 to 1000 Ohms for use with transcutaneous pacers.
Netech's EXPMT 2000 is menu driven and simple to use. It is compact light weight and cost half of the leading competetions model. It comes with customized hard carying case along with universal interface cables. Optional interface cables for various transcutanious cables are availble. Please check the "Ordering " information for part numbers or contact factory with specific manufactures models.
Show More...
Transthoracic Pacer
|
Test Load: 50 Ohm fixed, 12 selectable steps from 100 to 1000 ohms with EXPMT Load Selector |
Current: 0.1 to 25 mA |
Accuracy: +/- 10%, 1 to 10 mA +/- 5%, 10 to 200 mA |
Pulse Rate: 30 to 800 PPM |
Accuracy: +/- 1% or 2 ppm |
Pulse Width: 0.5 to 80 ms |
Accuracy: +/- 10%, 0.5 to 5 ms +/- 5%, 5 to 25 ms +/- 1%, 25 to 80 ms |
Sensed and Paced Refactory
|
Period: 25 to 500 ms |
Accuracy: +/- 10 ms |
0/60 Hz Immunity Test: 0 to 10.6 mV in 0.7 mV steps |
Transvenous Pacer
|
Test Load: 500 Ohms +/- 1% |
Accuracy: +/- 10%, 0.1 to 10 mA +/- 5%, 1 to 25 mA |
Pulse Rate: 30 to 800 PPM |
Accuracy: +/- 1% or 2 ppm |
Pulse Width: 0.5 to 80 ms |
Accuracy: +/- 10%, 0.5 to 5 ms +/- 5%, 5 to 25 ms +/- 1%, 25 to 80 ms |
Demand Sensitivity
|
Waveforms: SQR, TRI, and SSQ |
Amplitude: 0 to 100 mV |
Width: 10, 25, 40, 100, and 200 ms |
Pacer Rate: 30 to 120 ppm |
A - V Interval
|
0 to 999 ms |
General
|
Display: 4 line LCD display |
Housing: ABS plastic |
Power Requirements: 9V battery or AC operation |
Size: 7.75" x 4.0" x 1.5" (19.7 x 10.2 x 3.8 cm) |
Weight: 1 lb (0.9 kg) |
Environmental Operating Range: 59° to 95° F (15° to 35° C) |
Storage Temperature: 32° to 131° F (0° to 55° C) |
Part Number
|
Description
|
Price
|
950 |
_EXPMT 2000 External Pacemaker Analyzer with Case |
|
903 |
RS 232 interface cable, D9F-D9M |
|
950-LOAD-BOX |
Load Box for EXPMT 2000 |
|
905 |
Adapter cable for Zoll AED Pacemakers |
|
907 |
Adapter cable for Philips Medical AED Pacemakers |
|
906 |
Adapter cable for Medtronic AED Pacemakers |
|
908 |
Adapter Cable Banana to Straight Pin for Transvenous Pacemakers |
|
302 |
110/220V AC adapter |
|