Spo2 Simulator
OxSim - SpO2 Pulse Oximetry Tester
Product Code : VOX-1
The Netech Pulse OxSim SpO2 Oximetry Tester is the perfect complement to the SimCube family. It optically tests the entire system, including the sensor. Pocket sized, the OxSim runs on one AA alkaline battery for 10 hours. Test 4 Oxygen saturations, 3 pulse rates and a low perfusion mode all at the touch of a button. An indispensable tool for any biomed, OxSim fits in the SimCube carrying case and can be run simultaneously off the same battery boost option as your SimCube.
Show More...
Simulation Accuracy
|
Saturation Error: Error in the displayed saturation value should be less than or equal to the specified error of the sensor which is typically +/- 2%, but which can be +/-3% or more for some disposable sensors. |
Heart Rate Error: Less than 1 bpm |
Modes of Operation
|
85%: Saturation 85%; Pulse Rate 80 bpm ; Perfusion Index 2.0 |
95%: Saturation 95%; Pulse Rate 40 bpm; Perfusion Index 2.0 |
98%: Saturation 98%; Pulse Rate 80 bpm; Perfusion Index 2.0 |
140 BPM: Saturation 98%; Pulse Rate 140 bpm; Perfusion Index 2.0 |
Low Perfusion: Saturation 99%; Pulse Rate 80 bpm; Perfusion Index 0.2 |
Sensor Type Selection
|
Nell/Mas Used for: All Nellcor, Masimo, Datascope, Protocol, Spacelabs, HP/Philips (blue extension cable), HP/Philips (tan soft sensor), Nihon Kohden, Colin, GE/Marquette, Welch Allyn |
HP/Ohm Used for: HP cocoa brown soft sensor, Ohmeda and Novametrix |
General
|
Size: 2.5'' x 2.5'' x 1.25'' |
Weight: 5 ounces (141.75 g) excluding battery |
SPO2 Connection: Optical |
Power Battery: 1 AA battery |
Power Battery Life (Alkaline): 8 to 10 hours of simulation time (typical, depending on the oximeters under test); 60 days in sleep mode (unit sleeps after 10 min. with no probe connection) |
AC Operation: Included 6V DC supply operates over 100-240 VAC, 50-60 Hz |
 Brochure not available...
|
 Manual not available...
|
Part Number
|
Description
|
Price
|
OX-1 |
Optical SPO2 Simulator |
|