Basic Electrical Safety Analyzer
A Basic Electrical Safety Analyzer designed to test all Non-ECG equipment.
LKG 601
- A Comprehensive Electrical Safety Analyzer.
The LKG 601, Electrical Safety Analyzer is an extremely easy to use portable Electrical Safety Analyzer designed specifically for testing non-ECG equipment. Ground resistance, leakage current, and instrument current test measurements are displayed on its large LCD.
With its unique operating features, the LKG 601, Electrical Safety Analyzer
out performs all other basic electrical safety analyzers. To insure accuracy, each test has calibrated test points. The accessible instrument fuse solves the maintenance problems of built-in fuses. The selectable test load provides testing flexibility.
For the routine electrical safety testing of all your electrical equipment, the LKG 601,
Electrical Safety Analyzer is simply the best selection.
Features:
- Easy to Use
- Large LCD Display
- Accessible Line Fuse
- Calibrated Test Points
- Selectable Test Loads
- 15 and 20 Amp Capability
- Point-to-Point Testing
- Use any Test Cable
- 10 Lead Universal Connector
- Complies with AAMI/ESI 1993
- LED Unit of Measure Indicator
Measurements:
- Line Voltage
- Cord Resistance
- Case Leakage Current
- Earth Leakage Current
- Leads Leakage Current
- Leads ISO Test
- Instrument Current
- Point-to-Point Measurement
Back to top
Specifications:
Technical
Leakage Current:-
0-1999 micro amps. All Current measurements are made through
the AAMI/ANSI ESI - 1985 or IEC 601 - 1 test load. The meter readings correspond to the true RMS value of the current.
Accuracy: ±1% of reading, ±1 LSD DC to 1 kHz,
±2.5% of reading, ±1 LSD 1 kHz to 100 kHz,
±5% of reading, ±1 LSD 100 kHz to 1 mHz.
Resistance: 0 to 1999 milli ohms.
Accuracy is ±1% of reading, ±1 LSD
Device Current : 0 to 19.99 amps.
Accuracy is ±2% of FS, ±1LSD
Measures:-
Ground Resistance, Internal and External Leakage Current, and Instrument Current.
Test Receptacle:-
Hospital Grade 110V AC - 15 amp or 220V AC 10 amp. The rocker polarity switch selects NORMAL, OFF and REVERSE Polarity to the test receptacle, and a momentary Neutral switch will open the neutral line to the test receptacle.
General
Display: 3½ Digit LCD Display
Power Requirements: 110V AC 50 - 60 Hz, 15 amps
220V AC, 50 - 60 Hz, 10 amps
Physical Dimensions: 5.5 x 3.25 x 2.5 inches
(13.75 x 8.13 x 6.25 cms)
Weight: 1 lbs (0.45 kg)
Ambient Temperature:-
Operating Range: 59° to 95° F (15° to 40° C)
Storage Temperature: 0° to 122° F (-20° to 60° C)
Relative Humidity: 90% (max) at temperatures