Pacemaker Analyzers
DELTA 3000 - New Defib Analyzer /Pacer Tester
|
DELTA
3000, Defibrillator Analyzer is the latest addition of our
Defibrillator Analyzer with full Pacer Tester and ECG Simulator built
in.
DELTA 3000, Defibrillator Analyzer combines the complete
testing functions of a Defibrillator Tester, a Pacemaker Analyzer, and
a 12 Lead ECG / Arrhythmia simulator into one compact, lightweight, and
easy to use instrument. No other instrument delivers the performance
testing power of the DELTA 3000, Defib Analyzer / Pacer Tester.
DELTA
3000, Defibrillator Analyzer tests all defibrillators including AEDs.
It measures both monophasic and bi-phasic waveforms. It tests all
external pacemakers including transthoracic, transvenous, and AV
sequential models. It has 12 built-in user selectable test loads for
transthoracic pacing testing. The built-in ECG / Arrhythmia simulator
allows performance testing of any ECG equipment.
|
|
|
 |
 |
 |
Compact, Lightweight, and Rugged |
 |
Simple, Menu Driven Operation |
 |
Large 4-line LCD |
 |
Simultaneous Display of All Test Measurements |
 |
Dedicated Print Key for Immediate Documentaion |
|
 |
 |
RS 232 Interface |
 |
Measures: Energy, Peak Voltage, Peak Current, and Cardio Sync. Time |
 |
Battery or AC Operation |
 |
Storage and Playback of output waveforms |
 |
Two-Year Warranty |
 |
Tests all External Pacemakers - Transthorasic, Transvenous, and AV Pacers |
 |
Measures: Amplitude, Rate, Pulse Width, Pulse Energy, Sensed and Paced Refactory Periods |
 |
50 and 60 Hz Immunity Tests |
 |
Transthorasic Test Loads: Fixed 50 Ohm or 12 User Selectable Loads to 1000 Ohms |
 |
A / V Pacer Test Load: 500 Ohm |
 |
Tests all ECG Equipment |
 |
12 lead ECG, User Selectabel Rates and Amplitudes |
 |
Performance Waveforms: Sine, Square, Triangle, and Pulse |
 |
12 Arrhythmias |
|

|
|
Defibrillator Analyzer
Energy Measurement:
Load: 50 Ohm +/- 1 %
Range (High): 900 J Resolution: 1 Joule
Range (Low): 0 - 100 Resolution: 0.1 Joules
Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range
Maximum Voltage: 5200 V
Maximum Current: 100 Amp (H)
Sync. Time: 0 - 250 ms
Waveform Output:
Real Time: Output 1000 : 1 Stored.
Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output
Pacemaker Analyzer
Transcutaneous:
Test Load: 50 Ohm (Fixed) and variable 100 to 1000 Ohms user selectable
Current: 1 to 200 mA Accuracy: +/- 5% + 0.5 mA
Pulse Width: 0.5 to 80 ms Accuracy: +/- 1% + 0.1 ms
Refractory: 20 - 500 ms Accuracy: +/- 2% or + 10 ms
Atrial and Ventricular:
Test Load: 500 Ohms
Current: 1 to 25 mA Accuracy: +/- 5% + 0.5 mA
Pulse Rate: 30 to 800 ppm. Accuracy: +/- 1% + 0.1 ms
Refactory: +/- 20 - 500 ms
Accuracy: +/- 2% or 10 ms
Immunity Test: 50 / 60 Hz
ECG ArrhythmiaSimulator
ECG: Normal Sinus Rhythm.
Rate: 30, 60, 90, 120, 240
Performance Waveforms:
Sine, Square, Triangle and Pulse
Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz
Arrhythmias:
VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR
Waveform Output: 12 Lead ECG and Paddle and High Level Output Jacks.

|
Power Requirements:
Two 9V Alkaline Batteries Or AC Adapter
 
Dimension:
10 x 11 x 2.25 inches (25.4 x 27.9 x 5.7 cm)
Weight:
4 lbs (1.8 kg)
|
|
|
 |
|

 |