Transthoracic Pacer
Test Load: 50 Ohm fixed 12 Selectable steps from 100 to 1000 ohms with EXPMT Load Selector
Current: 0.1 to 25 mA
Accuracy:
+/- 10%, 1 to 10 mA
+/- 5%, 10 to 200 mA
Pulse Rate: 30 to 800 PPM
Accuracy: +/- 1% or 2 ppm
Pulse Width: 0.5 to 80 ms
Accuracy:
+/- 10%, 0.5 to 5 ms
+/- 5%, 5 to 25 ms
+/- 1%, 25 to 80 ms
Sensed and Paced Refractory
Period: 25 to 500 ms
Accuracy: +/- 10 ms
50/60 Hz Immunity Test:
0 to 10.6 mV in 0.7 mV steps
Transvenous Pacer
Test Load: 500 Ohms +/- 1%
Current: 0.1 to 25 mA
Accuracy:
+/- 10%, 0.1 to 10 mA
+/- 5%, 1 to 25 mA
Pulse Rate: 30 to 800 PPM
Accuracy: +/- 1% or 2 ppm
Pulse Width: 0.5 to 80 ms
Accuracy:
+/- 10%, 0.5 to 5 ms
+/- 5%, 5 to 25 ms
+/- 1%, 25 to 80 ms
Sensed and Paced Refractory
Period: 25 to 500 ms
Accuracy: +/- 10 ms
50/60 Hz Immunity Test:
0 to 102 mV in 6.8 mV steps
Demand Sensitivity
Waveforms: SQR, TRI, and SSQ
Amplitude: 0 to 100 mV
Width: 10, 25, 40, 100, and 200 ms
Pacer Rate: 30 to 120 ppm
A - V Interval
0 to 999 ms
|