Patient Simulators
MiniSim 1000 - Advanced - Multi-Parameter Arrhythmia Simulator
 |
The
MiniSim 1000, Arrhythmia Simulator is a powerful, comprehensive patient
simulator in a compact case. It is designed for testing the performance
of patient monitoring instrumentation quickly and easily.
The
microprocessor based instrument is menu driven via its tactile feel
keypad and easy to follow LCD display. The MiniSim 1000 provides full
12 lead ECG simulation with 14 user selectabel rates from 30 to 350 BPM
and 14 user selectable amplitudes from 0.15 to 5 mV. Sine, Square, and
Triangle performance waveforms are simulated with 14 user selectable
rates from 0.1 to 100 Hz.
The MiniSim also simulates 45
different arrhythmias including atrial, ventricular, tall T wave, ST
elevation, ST depression, myocardial infraction, blocks, pulse, and
pacer waveforms. In addition, two invasive blood pressures are
simulated with choice of 12 static pressures and 6 dynamic waveforms.
Respiration is simulated with 4 rates. 4 base line impedance, and 4
delta impedance selections. Apnea may be off, continuous, and with
either 12 or 32 second apnea bradycardia. Finally, temperature is
available at either 25 or 37 degrees Centigrade.
The MiniSim
1000 multi-parameter patient simulator is powered by a 9 volt alkaline
battery with a typical life of 40 continuos hours. The instrument
provides unmatched choices of parameter simulation in a truly compact
instrument.
|
|
|
 |
 |
 |
12 Lead ECG |
 |
45 Arrhythmias |
 |
ECG Performance Waveforms |
 |
2 Blood Pressure |
 |
Respiration / Apnea |
 |
Temperature |
 |
Easy to Use |
 |
Small, Compact Size |
 |
Menu driven operatoin |
 |
Large LCD Display |
 |
Tactile Keypad |
 |
Battery or AC Operation |
|

|
|
ECG: 12 leads with independent outputs referenced to RL.
NORMAL SINUS RHYTHM (NSR)
ECG Rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. Accuracy 0.5%.
Amplitudes:
0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is
0.6 X Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as
Lead II.
High Level: 500 X low level output on Lead II.
Accuracy: 2% (1-5mV).
PERFORMANCE WAVEFORMS (PERF)
Sine, Square, Triangle, Pulse
Frequencies: 0.1 to 0.9 in 0.1 Hz steps.
1.0 to 9.0 in 1.0 Hz steps.
10 to 100 in 10 Hz steps.
Accuracy: 1%
Amplitudes: 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II.
Pulse: 20 ms pulse of 1mV amplitude repeated at 4 second intervals.
R WAVE DETECTION (RWD)
Widths: 10, 40, 50, 60, 70, 80, 90, 100, 110, and 120 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 mV on Lead II.
PACER (PCR)
AP: Atrial Pacer set at 70 BPM.
ASP: Asynchronous Pacer � Ineffective Pacing.
Pacer Amplitudes: + 2 mV and �2mV.
QRS Amplitudes: 1 mV
PCR: Pacer pulses alone.
Pacer Widths: 0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV.
VP: Ventricular Pacer set at 70 BPM.
AVP: Atrial Ventricular Pacer set at 70 BPM.
QRS Amplitude: 1 mV.
Pacer Amplitude: -2mV.
RESPIRATION
Baseline Impedance: 250, 500, 750, and 1000 Ohms.
Delta Impedance: 0.1, 0.5, 1.0, and 1.5 Ohms.
Accuracy: 10%.
Rates: 15, 30, 60, and 120 BPM.
Accuracy: 1%
Apnea: Off, continuous, 12 seconds and 32 seconds.
Lead Configuration: Leads I and II.
BLOOD PRESSURE
Impedance: 350 Ohms.
Excitation: 2 to 16 Volts.
Sensitivity: 5mV/V/mmHg.
Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, and 300 mmHg.
Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, and 120/80 mmHg. Dynamic waveforms track all NSR rates.
TEMPERATURE
Compatibility: YSI 400 and 700 Series.
Temperature: 25 and 37 degrees Centigrade.
Accuracy: 2% of setting
|
POWER REQUIRMENTS:
One 9 Volt alkaline battery or optional AC adapter.
PHYSICAL CHARACTERISTICS:
Dimensions: 5.5 X 3.5 X 1.5 inches (13.9 X 8.9 X 3.8 cm).
Weight: 10 oz ( 0.3 kg).
TEMPERATURE REQUIREMENTS:
Operating: 59 to 95� F (15 to 35�C).
Storage: 32 to 131� F (0 to 55�C).
|
|
|
 |
|

 |